A Random Access, Fully-Automated, "Walk Away" Clinical Chemistry Analyzer with a throughput of up to 100 tests per hour (up to 150 tests per hour with ISE).
![]() |
Throughput
- Up to 100 tests per hour;
- Up to 150 tests per hour with ISE unit.
Reagent & Sample system up↑
One rotor combining both sample and reagent positions;
Inner rotor ring:
-15 positions for reagent bottles
-Optionally cooled to approx. 12°C below ambient temperature;Outer rotor ring:
- 25 positions for samples, controls and calibrators;
- 15 positions for reagent bottles at ambient temperature;All reagent positions can be assigned as R1, R2 and R3;
Reagent 1 volume 110 - 400 µl;
Reagent 2 volume 0 - 180 µl;
Reagent 3 volume 0 - 180 µl;
Preheated reagent and sample needle with level detection and integrated mixer;
Typical reagent consumption 250 µl per test;
Continuous sample loading;
13 mm OD primary tubes;
All sample positions can contain 5 ml primary tubes or sample cups;
Sample volume 1 - 30 µl per test, programmable in steps of 0.1 µl.
Sample predilution
- Programmable ratios 1:5, 1:10, 1:20, 1:30, 1:40, 1:50 with 3 possible diluents.
Cuvette rotor up↑
- Semi-disposable rotor with 48 cuvettes. Path length 7 mm;
- > 10,000 tests per rotor;
- Minimum measuring volume 220 µl;
- Measuring temperature 37°C, controlled by Peltier elements;
- Cuvette washing and drying.
Light source
- Quartz-iodine lamp 12V-20W.
Wavelength range
- Automatic wavelength selection by 8-position filter wheel (340, 376, 405, 436, 505, 546, 578 and 620 nm);
- Other wavelengths available on request.
Photometric range
- -0.1 to 3.0 Absorbance.
Analytical modes up↑
- Kinetic measurement with linearity check;
- Bichromatic end point measurement with or without bichromatic reagent blank and/or sample blank correction;
- Two point measurement;
- Graphic plot of all measuring points;
- Automatic rerun with sample reduction;
- Non-linear calibration curves.
Calculation modes
- Prozone check for immunology tests;
- Cut-off declaration.
Quality control
- Up to 15 different controls can be defined, 3 per test;
- Westgard rules;
- Levey-Jennings plots.
Water Consumption up↑
- 0.5 liter per hour.
Standards
- CE, CB certificate, UL
Internal Computer
- Pentium processor;
- LCD Monitor;
- Keyboard, mouse;
- MS Windows XP operating system.
Dimensions (Benchtop)
- 89 x 58 x 60 cm (W x H excl. Monitor x D);
- No peripherals under lab table.
Options
- Bar code reader:
- Hand held CCD bar code reader (can read all common barcodes) used for test requisition and reagent identification.- ISE Unit:
- Na, K and Cl measurements.
- CO2 measurement optional.
|
|
||||
|
|